Máy đo độ dẫn điện-EC, chất rắn hòa tan-TDS, đo độ muối, Nhiệt độ dạng cầm tay

HI-98360 EC/TDS/NaCl/°C Meter

Hãng sản xuất: HANNA

Xuất xứ: Romania

Model: HI 98360

máy đo NaCl cầm tay Hanna (http://thietbidohanna.com/may-do-muoi-do-man-hanna-128.html)

EC/TDS/NaCl/°C meter with PC connectivity and logging, Ranges: EC 0.00-500.0 mS/cm, TDS 0.00-400.0 g/L, NaCl 0.0-400.0%, Temp -9.9 to 120.0°C

Giá: Liên hệ 0977.459.400

Rugged all in one conductivity meter with datalogging

Rugged and highly versatile, this high precision instrument combines the four parameters needed to measure conductivity in one high precision multiparameter instrument.

Designed to allow simultaneous sampling of electro conductivity (EC), total dissolved solids (TDS), salinity (NaCl) and temperature, this portable meter is ideal for measuring conductivity in testing or pressured environments.

With a rugged waterproof casing, they are ideal for use out in the field as well as internal laboratory testing.

Four ring probe design for added stability over broader measuring range

The HI 76309 conductivity and temperature probe features a PVC body with a stainless steel four ring design to deliver high accuracy over a broader conductivity range.

Hanna HI 98360 (http://thietbidohanna.com/may-do-muoi-do-man-hanna/may-do-ec-tds-nacl-nhiet-do-cam-tay-hanna-hi-98360-664.html)

The four ring construction is designed to ensure the probe is immune to polarisation and fouling for longer periods of time.

Combined with either a 1m or 1.5m cable, this probe makes the meter particularly popular for use out in the field.

Key features
Water resistant, rugged ergonomic casing - designed to survive harsh conditions of field use Data logging up to 500 records with USB port featured as standard Large backlit LCD display Clear tutorial messaging guides the user through calibration, operation and setup Battery error prevention system alerts user if power is low % battery displayed on start up Rubberised keyboard with clearly marked buttons for intuitive use Auto range of EC and TDS modes automatically set meter to scale with highest possible resolution Auto end point automatically freezes the display once a stable reading is reached Selectable conductivity and TDS measurement Temperature coefficient from 0.00 to 6.00°C Temperature reference from 20 to 25°C Selectable TDS factor of 0.40 to 0.80 4-ring probe design Promotes greater stability over a broader measuring range Equipped with an embedded temperature sensor Immunity to polarisation and falling Probe connector located on top of instrument for ease and efficiency
Máy đo EC/TDS/NaCl/Nhiệt độ cầm tay Hanna HI 98360 (http://thietbidohanna.com/may-do-muoi-do-man-hanna/may-do-ec-tds-nacl-nhiet-do-cam-tay-hanna-hi-98360-664.html)
Added benefits
Data logging up to 500 records with USB port for data transfer included with HI 98360 Choose between automatic and manual temperature compensations to match operating requirements Meter can be connected to a 12 Vdc power supply (optional)

Supplied with
HI 76309/1.5 conductivity probe with internal temperature sensor on 1.5m cable Batteries Instructions Rugged carrying case


Range

EC

0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm;
3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm (actual EC)



Range

TDS

0.00 to 14.99 ppm (mg/L); 15.0 to 149.9 ppm (mg/L); 150 to 1499 ppm (mg/L);
1.50 to 14.99 g/L; 15.0 to 100.0 g/L up to 400.0 g/L (actual TDS)



Range

NaCl

0.0 to 400.0%



Range

Temperature

9.9 to 120.0°C



Resolution

EC

0.01 µS/cm; 0.1 µS/cm; 1 µS/cm;
0.01 mS/cm; 0.1 mS/cm



Resolution

TDS

0.01 ppm (mg/L); 0.1 ppm (mg/L); 1 ppm (mg/L);
0.01 g/L; 0.1 g/L



Resolution

NaCl

0.1%



Resolution

Temperature

0.1°C



Accuracy

EC

±0.5% of reading



Accuracy

TDS

±0.5% of reading



Accuracy

NaCl

±0.5% of reading



Accuracy

Temperature

±0.4°C



Calibration

EC

automatic, 1 point, with 6 memorized values
(84, 1413, 5000, 12880, 80000 and 111800 µS/cm)



Calibration

NaCl

1 point, with HI 7037 calibration solution (not included)



Calibration

Temperature

2 point, at 0 and 50°C (32 to 122°F)



Temperature Compensation



automatic or manual, 0 to 60°C (32 to 140°F);
can be disable for measuring actual EC and TDS



Temperature Coefficient



adjustable from 0.00 to 6.00%/°C (EC and TDS only)



TDS Conversion Factor



adjustable from 0.40 to 0.80



Data Logging



up to 250 samples



Probe



HI 76309/1.5, with 1.5 m (4.9') cable (included)



Power Supply



4 x 1.5V AA batteries (included) or 12 Vdc input;
auto-off after 5 minutes of non-use (can be disabled)



Environment



0 to 50°C (32 to 122°F); RH max 100%



Dimensions



196 x 80 x60 mm (7.7 x 3.1 x 2.4")



Weight



500 g (1.1 lb.)